it can be calculated that a 1 mV error in any of the potential terms corresponds to an error of about 4% in the hydrogen-ion activity. Under carefully controlled experimental conditions, the potential of a pH cell can be measured with an uncertainty as small as 0.3 mV, which corresponds to a ± 0.005 pH unit uncertainty.

The measurement of pH using the operational cell assumes that no residual liquid-junction potential is present when a standard buffer is compared with a solution of unknown pH. Although this may never strictly be true, especially with complex matrices, the residual liquid-junction potential can be minimized by the appropriate choice of a salt-bridge solution and calibration buffer solutions.

pH MEASUREMENT SYSTEM ELECTRODES Glass Electrodes

The glass electrode is the hydrogen-ion sensor in most pH-measurement systems. The pH-responsive surface of the glass electrode consists of a thin membrane formed from a special glass that, after suitable conditioning, develops a surface potential that is an accurate index of the acidity of the solution in which the electrode is immersed. To permit changes in the potential of the active surface of the glass membrane to be measured, an inner reference electrode of constant potential is placed in the internal compartment of the glass membrane. The inner reference compartment contains a solution that has a stable hydrogen-ion concentration and counter ions to which the inner electrode is reversible. The choice of the inner cell components has a bearing on the temperature coefficient of the emf of the pH assembly. The inner cell commonly consists of a silver-silver chloride electrode or calomel electrode in a buffered chloride solution.

Immersion electrodes are the most common glass electrodes. These are roughly cylindrical and consist of a barrel or stem of inert glass that is sealed at the lower end to a tip, which is often hemispherical, of special pH-responsive glass. The tip is completely immersed in the solution during measurements. Miniature and microelectrodes are also commercially available and used widely, particularly in physiological studies. Capillary electrodes permit the use of small samples and provide protection from exposure to air during the measurements, for example, for the determination of blood pH. This type of electrode may be provided with a water jacket for temperature control.

The membrane of pH-responsive glass usually is made as thin as is consistent with adequate mechanical strength; nevertheless, its electrical resistance is high, for example, 10 to 250 Mi2. Therefore, an electronic amplifier must be used to obtain adequate accuracy in the measurement of the surface potential of a glass electrode. The versatility of the glass electrode results from its mechanism of operation, which is one of proton exchange rather than electron transfer; hence, oxidizing and reducing agents in the solution do not affect the pH response.

Most modern electrode glasses contain mixtures of silicon dioxide, either sodium or lithium oxide, and either calcium, barium, cesium, or lanthanum oxide. The latter oxides are added to reduce spurious response to alkali metal ions in high pH solutions. The composition of the glass has a profound effect on the electrical resistance, the chemical durability of the pH-sensitive surface, and the accuracy of the pH response in alkaline solutions. Both the electrical and the chemical resistance of the electrode glasses decrease rapidly with a rise in temperature. Therefore, it is difficult to design an electrode that is sufficiently durable for extended use at high temperatures and yet, when used at room temperature, free from the sluggish response often characteristic of pH cells of excessively high resistance. Most manufacturers use different glass compositions for electrodes, depending on their intended use.

The mechanism of the glass electrode response is not entirely understood. It is clear, however, that when a freshly blown membrane of pH-responsive glass is first conditioned in water, the sodium or lithium ions that occupy the interstices of the silicon-oxygen network in the glass surface are exchanged for protons from the water. The protons find stable sites in the conditioned gel layer of the glass surface. Exchange of the labile protons between these sites and the solution phase appears to be the mechanism by which the surface potential reflects changes in the hydrogen ion activity of the external solution. When the glass electrode and the hydrogen gas electrode are immersed in the same solution, their potentials usually differ by a constant amount, even though the pH of the medium is raised from 1 to 10 or greater. In this range, the potential Ee of a glass electrode may be written

Sleeping Sanctuary

Sleeping Sanctuary

Salvation For The Sleep Deprived The Ultimate Guide To Sleeping, Napping, Resting And  Restoring Your Energy. Of the many things that we do just instinctively and do not give much  of a thought to, sleep is probably the most prominent one. Most of us sleep only because we have to. We sleep because we cannot stay awake all 24 hours in the day.